Control of postpartum hemorrhage using vacuum-induced uterine tamponade

Yuditiya Purwosunu, Widyastuti Sarkoen, Sabaratnam Arulkumaran, Jan Segnitz

Research output: Contribution to journalArticlepeer-review

Abstract

BACKGROUND: Postpartum hemorrhage is the leading cause of maternal mortality worldwide. Vacuum-induced uterine tamponade is a possible alternative approach to balloon tamponade systems for the treatment of postpartum hemorrhage resulting from atony. METHOD: In a prospective proof-of-concept investigation of 10 women with vaginal deliveries in a hospital setting who failed first-line therapies for postpartum hemorrhage, tamponade was used. Vacuum-induced uterine tamponade was created through a device inserted transvaginally into the uterine cavity. An occlusion balloon built into the device shaft was inflated at the level of the external cervical os to create a uterine seal. Negative pressure was created by attaching a self-contained, mobile, electrically powered, pressure-regulated vacuum pump with a sterile graduated canister. EXPERIENCE: In all 10 cases, the suction created an immediate seal at the cervical os, 50-250 mL of residual blood was evacuated from the uterine cavity, the uterus collapsed and regained tone within minutes, and hemorrhaging was controlled. The device remained in place for a minimum of 1 hour and up to 6.5 hours in one case while vaginal and perineal lacerations were easily repaired. CONCLUSION: This preliminary investigation suggests that a device designed to create vacuum-induced uterine tamponade may be a reasonable alternative to other devices used to treat atonic postpartum hemorrhage.

Original languageEnglish
Pages (from-to)33-36
Number of pages4
JournalObstetrics and Gynecology
Volume128
Issue number1
DOIs
Publication statusPublished - 1 Jul 2016

Fingerprint

Dive into the research topics of 'Control of postpartum hemorrhage using vacuum-induced uterine tamponade'. Together they form a unique fingerprint.

Cite this