Validation of traction force indicator in ventouse devices

B. J. Whitlow, O. Tamizian, J. Ashworth, S. Kerry, L. K. Penna, S. Arulkumaran

Research output: Contribution to journalArticlepeer-review

Abstract

Objective: To evaluate whether the traction force indicator in a ventouse device gave an accurate estimation of the force applied. Methods: The study was carried out at the Derby City Hospital. An estimate of measurement errors was made by analyzing the standard deviation of the residuals for 30 Kiwi OmniCup devices (Clinical Innovations, Abingdon, Oxfordshire, England) and a standard industrial spring balance. Results: The measurement errors for the traction forces were very small and acceptable for the OmniCup. The overall estimate of measurement errors was 0.45 kg (1.35 lb). Conclusion: The overall measurement error for the traction force indicator of the OmniCup was found to be very small. Injuries associated the ventouse, such as cephalohematomas and subgaleal hemorrhages, may be minimized if the recommended limits for a safe traction force are not exceeded.

Original languageEnglish
Pages (from-to)35-38
Number of pages4
JournalInternational Journal of Gynecology and Obstetrics
Volume90
Issue number1
DOIs
Publication statusPublished - 2005